Photonics Spectra
| Accent on Applications | May 2006
Deformable
Mirror Improves Microscopes Field of View
by Hank Hogan
Benjamin Potsaid, a research scientist at the
Center for Automation Technologies and Systems
at Rensselaer Polytechnic Institute in Troy, N.Y.,
wants to view small objects over long distances.
Others at the center who are working on robotic
micromanipulation and assembly would like to magnify
small objects over large areas.
The problem is that physics and real-life optics
get in the way. As the resolution of the
microscope increases to observe smaller features,
the field of view the region of the object
that can be observed decreases, Potsaid
explained.
To get around this, the researchers turned to
deformable mirror technology from Boston Micromachines
Corp. of Watertown, Mass. They have constructed
an adaptive scanning optical microscope that employs
the deformable mirror to correct for off-axis
aberrations in a custom scanner lens. Potsaid
said that this new technology enables a field
of view that is two orders of magnitude larger
at the same resolution than a conventional microscopes,
citing a design with a 40-mm-diameter field of
view at a resolution of 1.5 µm.
There are several traditional solutions to the
problematic trade-off between resolution and field
of view. These include the use of multiple parfocal
objectives, zoom lenses, moving stages, moving
microscopes and multiple microscopes. It even
is possible to get close-ups over long distances
using a high-pixel-count camera and the optics
employed in projection lithography. The last approach
is costly, however, and the others suffer from
such drawbacks as agitating the specimen during
scanning, a limited field of view, slow scanning
rates or difficult integration with conveyor transport.
As their solution, the scientists at the center
had devised an instrument with a scanner lens
assembly and a steering mirror. The mirror would
move the field of view over a sample, and a camera
would capture a series of small, distinct snapshots.
These images then would be assembled into a large
mosaic. Simulations showed that the approach would
work provided that off-axis aberrations
in the optics were corrected.
Then, just as the team was transitioning from
simulation to prototype design and production,
the Mini-DM from Boston Micromachines became available.
A smaller version of a microelectromechanical
systems deformable mirror that the company had
been offering, the mirror has a 2-mm-diameter
aperture and 32 actuators that produce a 2.5-
to 3.5-µm stroke. The actuators are spaced
400 µm apart in a grid across the area of
the device. On command, the actuators move, thereby
changing the mirrored surface slightly at up to
1000 Hz.
The deformable mirror corrects
for off-axis aberrations in the adaptive scanning
optical microscope, yielding larger fields of
view at the same resolution than a conventional
microscope provides. Courtesy of Center for Automation
Technologies and Systems, Rensselaer Polytechnic
Institute.
That capability allows
the microscope to correct for off-axis optical
aberrations. When the steering mirror is angled
properly, the image at the camera comes from the
center line of the lens assembly and is, therefore
nearly perfect. In such a case, the Strehl ratio
a figure of merit that compares the performance
of actual and ideal optics would be 0.97
or 0.98 over the field of view. When the angle
of the steering mirror changes, the field of view
shifts. The images then come from an off-axis
point and are much worse than the ideal if not
corrected.
By adjusting the deformable
mirrors shape, the researchers account for
these effects, which are known in advance, thanks
to measurements and a calibration procedure. As
a result of putting the right hills and valleys
in the deformable mirror, the researchers keep
the adaptive scanning optical microscopes
Strehl ratio at 0.97 and higher over a 40-mm range
of travel.
In addition to offering
a larger field of view than is possible with traditional
instruments, the microscope enables scanning speeds
10 to 100 times higher than those achieved by
moving stages, Potsaid said. The microscope also
is time-efficient in low-light conditions, which
is important in biological studies where photo-bleaching
of fluorescent labels can occur. The pixels
are exposed simultaneously, rather than sequentially,
as is the case in most confocal or line-scan imaging
approaches, he said.
Additionally, the instrument offers different
modes of operation, such as imaging regions of
interest, tracking multiple moving objects and
full-area coverage. A twist to the latter is that
the microscope combines low-resolution background
monitoring for rare-event detection and high-resolution
imaging of the event itself.
There are some disadvantages to the approach,
however. For example, a moving stage offers a
virtually unlimited field of view, whereas the
adaptive scanning optical microscope does not.
On the other hand, because the adaptive scanning
optical microscopes stage does not move,
there is no chance that a delicate specimen will
be disturbed or agitated.
The technology could be used in robotic micromanipulation,
biotechnology and medicine, industrial quality
assurance and automated medical diagnostics. Because
of its wide applicability, Potsaid does not see
the adaptive scanning optical microscope remaining
exclusively at the center. He said that discussions
are under way regarding commercialization and
licensing of the intellectual property.
Contacts:
Benjamin Potsaid, Rensselaer Polytechnic Institute,
Troy, N.Y.; +1 (518) 276-8707
Paul Bierden, Boston
Micromachines Corp., Watertown, Mass.; +1 (617)
926-4178;
|
 |